Customization: | Available |
---|---|
Classification: | Physiological Functions of Diagnosis and Monitoring Equipment |
Type: | Digital Thermometer |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Specifications |
Measurement System : 8-Channel optical system Measurement Range : 0.000-4.000Abs Photometric Accuracy : ±1% or ±0.001Abs Repeatability : <±0.3% Stability : <0.002Abs Linearity : <±0.5% Reading Speed : <5secs/96holes(single wavelength), <7secs/96holes(double wavelength) Filters : Standard: 4 filters (405, 450, 492,630 nm) or other wavelength on request..Totally 10 filter can be installed. Full One year Warranty |
Accuracy of wavelength (nm) | Half-width(nm) |
±3.0 | ≤12 |
Absorbance (A) | Accuracy Error |
0.000-2.000 | ±0.01 |
>2.000-3.000 | ±0.05 |
>3.000-4.000 | ±0.15 |
Absorbance (A) | Error (%) |
0.000-2.000 | ±1.0 |
>2.000-3.000 | ±2.0 |
>3.000-4.000 | ±5.0% |
Technical Specification | ||||
ELISA microplate reader : | Type: HP-ELISA9600 | |||
1. Stand-alone microprocessor controlled photometer designed to measure absorbance in 96 well (12X8) microplates. | Yes |
|||
2. Programmable reading between 1 to 96 wells | Yes | |||
3. Reading time: single wave 5s/96 plate, dual wave 7s/96 plate; | Yes | |||
4. Reading Domain: up to 4.0 OD in range 400 - 850 nm | Yes | |||
5. Dynamic range: 0 to 4.000 OD | Yes | |||
6. Linearity: < ±0.5% from 0 - 3.500 OD | Yes | |||
7. Precision: 0 to 3.500 OD: ±1% or ±0.001 Abs CV max | Yes | |||
8. Wavelength range: 400-800 nm | Yes | |||
9. Light source: Tungsten halogen lamp | Yes | |||
10. Filters: up to 6 changeable filters (on board to be delivered with at least 450 nm, 490 nm, 620 nm) | Yes | |||
11. Capacity: 120 assay procedure, 100 plates, and 20 standard curve data (for quantitative determination) | Yes | |||
12. Result interpretation: Positive, negative, doubtful | Yes | |||
13. Plate validation: up to 6different criteria, possibility of exclusion of nonconforming values and reinterpretation of the results | Yes | |||
14. Results printing: exactly data, and Negative ,Positive | Yes | |||
15. Self test | Yes | |||
16. LCD display | Yes | |||
17. Power: 220-240 V 50/60 Hz | Yes | |||
18. Printer interface: (to be delivered with printer) | With Inner heat printer or external printer,all is ok | |||
19. Computer interface | Yes RS232 interface | |||
20 .If you need, we also can supply Laser, printer | Yes ( We can give a Laser printer for this machine ) |
|||
20.1.Compatible with "ELISA Microplate automated reader" | Yes | |||
20.2. A4 paper format | ||||
20.3. Spare cartridge -1 | Yes | |||
20.4. Printing speed not less then 3 ppm | Yes | |||
20.5. Compatible MS Windows | Yes | |||
20.6. - Workload - 1000 pages per month, Printer resource - 20 000 hours | Yes | |||
20.7. Power source -220 v | Yes | |||
General Conditions | ||||
1. The installation is so simple that operators can install it by themselves according to the user manual. 2. HEPO INTERNATIONAL will provide the customers with free training service in our factory.The traffic and accommodation charge will be on the customers' account. Also, HEPO INTERNATIONAL will provide free technical support via e-mail, fax or telephone at any moment. 3. Full warranty - 1 years 4. Technical documentation in English Full One year Warranty |