Customization: | Available |
---|---|
Classification: | Biological Diagnostics |
Type: | Urine Analyzer |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Model | Screen size (inch ) | Puncture | Storage | Graphical Touch Screen | Power |
HP-HEMA6000V | 10.4 | None | 20K | NO | 50Hz±3Hz |
Specifications | |||
Test Items | WBC, LY#, MO#, GR#, LY%, MO%, GR%, HGB, RBC, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PCT, MPV, PDW, WBC Histogram, RBC Histogram, PLT Histogram | ||
Background Count | WBC: ≤0.3×109/L RBC: ≤0.03×1012/L HGB: ≤1g/L PLT: ≤10×109/L | ||
Linearity | Test Item | Test Range | Allow Difference of Linearity |
WBC | 1.0∼10.0×109/L | ±0.5×109/L | |
10.0∼99.9×109/L | ±5% | ||
RBC | 0∼1.0×1012/L | ±0.05×1012/ L | |
1.0∼9.99×1012/L | ±5% | ||
HGB | 0∼99 g/L | ±2g/L | |
20∼70 g/L | ±3% | ||
PLT | 0∼100×109/L | ±10×109/L | |
101∼999×109/L | ±10% | ||
Repeatability | Test Item | CV(%) | Teat Range |
WBC | ≤2.0 | 7.0~15.0×109/L | |
RBC | ≤0.5 | 3.5~6.0×1012/L | |
HGB | ≤1.5 | 110~180g/L | |
PLT | ≤4.0 | 150~500×109/L | |
MCV | ≤0.5 | 80~110fL | |
Sample Volume | Twig blood: 13μl | ||
Dilution Ratio | WBC/HGB: 1:500 RBC/PLT: 1:50000 | ||
Interfaces | RS232 joint, USB, PS2, Parallel port | ||
Test Speed | 60 samples/hours | ||
Continuous Work Time | >8h | ||
Power Requirement | AC110-240V ,47HZ-63HZ |